Specifications
○
C
Copyright 2012 Agilent Technologies
Change 6
Revised Minimum, Maximum & Measurement Uncertainty in Impedance Measurement Accuracy Test
(100pF STANDARD (0 m)).
100 pF STANDARD (0 m) C.V.: (Cal.Value at 1MHz)
Test Frequency: 1MHz
Upper Value: Capacitance (Cp)
Lower Value: Dissipation Factor (D)
Test Signal Level
Minimum
Measured
Value
Maximum
Measurement
Uncertainty
5 mV
C.V. - 0.636 pF
C.V. + 0.636 pF
0.0250 pF
-0.0064
0.0064
0.0002
20 mV
C.V. - 0.222 pF
C.V. + 0.222 pF
0.0250 pF
-0.0022
0.0022
0.0002
100 mV
C.V. - 0.153 pF
C.V. + 0.153 pF
0.0250 pF
-0.0015
0.0015
0.0002
1 V
C.V. - 0.153 pF
C.V. + 0.153 pF
0.0250 pF
-0.0015
0.0015
0.0002
2 V
C.V. - 0.218 pF
C.V. + 0.218 pF
0.0250 pF
-0.0022
0.0022
0.0002