Specifications

C
Copyright 2012 Agilent Technologies
Change 5
Revised Minimum, Maximum & Measurement Uncertainty in Impedance Measurement Accuracy Test (10pF
STANDARD (0 m)).
10 pF STANDARD (0 m) C.V.: (Cal.Value at 1MHz)
Test Frequency: 1MHz
Upper Value: Capacitance (Cp)
Lower Value: Dissipation Factor (D)
Test Signal Level
Minimum
Measured
Value
Maximum
Measurement
Uncertainty
5 mV
C.V. - 0.0689 pF
C.V. + 0.0689 pF
0.0025 pF
-0.0069
0.0069
0.0002
20 mV
C.V. - 0.0307 pF
C.V. + 0.0307 pF
0.0025 pF
-0.0031
0.0031
0.0002
100 mV
C.V. - 0.0244 pF
C.V. + 0.0244 pF
0.0025 pF
-0.0024
0.0024
0.0002
1 V
C.V. - 0.0244 pF
C.V. + 0.0244 pF
0.0025 pF
-0.0024
0.0024
0.0002
2 V
C.V. - 0.0212 pF
C.V. + 0.0212 pF
0.0025 pF
-0.0021
0.0021
0.0002