Specifications

C
Copyright 2012 Agilent Technologies
Change 3
Revised Minimum, Maximum & Measurement Uncertainty in Test Signal Level Accuracy Test.
Test Frequency
Test Signal
Level
Monitor
Reading
Maximum
Measurement
Uncertainty
100 kHz
1 V
1.081 V
0.042 V
1 MHz
5 mV
6.420 mV
0.210 mV
1 MHz
20 mV
22.68 mV
0.840 mV
1 MHz
100 mV
109.4 mV
4.200 mV
1 MHz
1 V
1.085 V
0.042 V
1 MHz
2 V
2.169 V
0.084 V
10 MHz
1 V
1.121 V
0.060 V
30 MHz
1 V
1.201 V
0.100 V
30 MHz
2 V
2.401 V
0.200 V
Change 4
Revised Minimum, Maximum & Measurement Uncertainty in Impedance Measurement Accuracy Test (1pF
STANDARD (0 m)).
IMPEDANCE MEASUREMENT ACCURACY TEST
1 pF STANDARD (0 m) C.V.: (Cal.Value at 1MHz)
Test Frequency: 1MHz
Upper Value: Capacitance (Cp)
Lower Value: Dissipation Factor (D)
Test Signal Level
Minimum
Measured
Value
Maximum
Measurement
Uncertainty
5 mV
C.V. - 0.04000 pF
C.V. + 0.04000 pF
0.0005 pF
-0.0400
0.0400
0.0002
20 mV
C.V. - 0.01135 pF
C.V. + 0.01135 pF
0.0005 pF
-0.0114
0.0114
0.0002
100 mV
C.V. - 0.00371 pF
C.V. + 0.00371 pF
0.0005 pF
-0.0037
0.0037
0.0002
1 V
C.V. - 0.00275 pF
C.V. + 0.00275 pF
0.0005 pF
-0.0028
0.0028
0.0002
2 V
C.V. - 0.00228 pF
C.V. + 0.00228 pF
0.0005 pF
-0.0023
0.0023
0.0002