Agilent 16334A Test Fixture Operation and Service Manual Second Edition Agilent Part No.
Notices The information contained in this document is subject to change without notice. This document contains proprietary information that is protected by copyright.All rights are reserved. No part of this document may be photocopied, reproduced, or translated to another language without the prior written consent of the Agilent Technologies. Agilent Technologies Japan, Ltd. Component Test PGU-Kobe 1-3-2, Murotani, Nishi-Ku, Kobe-shi, Hyogo, 651-2241 Japan © Copyright Agilent Technologies Japan, Ltd.
Operating personnel must not remove instrument covers. Component replacement and internal adjustments must be made by qualified maintenance personnel. Do not replace components with the power cable connected. Under certain conditions, dangerous voltages may exist even with the power cable removed. To avoid injuries, always disconnect power and discharge circuits before touching them.
CAUTION This Caution sign denotes a hazard. It calls attention to a procedure, practice, condition or the like, which, if not correctly performed or adhered to, could result in damage to or destruction of part or all of the product. NOTE Note denotes important information. It calls attention to a procedure, practice, condition or the like, which is essential to highlight.
IMPORTANT No other warranty is expressed or implied. Agilent Technologies specifically disclaims the implied warranties of merchantability and fitness for a particular purpose. Exclusive Remedies The remedies provided herein are buyer’s sole and exclusive remedies. Agilent Technologies shall not be liable for any direct, indirect, special, incidental, or consequential damages, whether based on contract, tort, or any other legal theory.
Contents 1. Operation DESCRIPTION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Typical Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Contents 8
1 Operation This operation note provides complete information on the 16334A Test Fixture.
Operation DESCRIPTION DESCRIPTION The 16334A Testfixture was designed for use with the following instruments: 4192A LF lmpedance Analyzer 4194A lmpedanoe/Gain-Phase Analyzer 4274A Mu-ti-Frequency LCR Meter 4275A Multi-Frequency LCR Meter 4276A LCZ Meter 4277A LCZ Meter 4278A 1kHz/1MHz Capacitance Meter The 16334A is a direct attachment, 4-terminal pair type fixture which is equipped with a tweezer type probe for measurement of leadless components such as chip capacitors. The tweezer tips are replaceable.
Operation DESCRIPTION Table 1-1 Accessories Description Agilent Part No Tweezer Tip 16334•09002 (1ea.
Operation Specifications Specifications Table 1-2 Function: For use with 4192A, 4194A, 4274A, 4275A, 4276A, 4277A and 4278A instruments. Permits connecting miniature leadless components to the UNKNOWN terminals (4-terminal pair configuration) of the 4192A, 4194A, 4274A, 4275A, 4276A, 4277A, or 4278A. Measurable Components: Miniature components with terminals on opposite faces (Distance between component terminals ≤ 10mm).
Operation Typical Characteristics Typical Characteristics Table 1-3 Typical Characteristics Applicable Measurement Range Incremental error (1 MHz≤f ) Parameter Value Measurement Frequency Parameter reading error (%) Offset value for D 4192A Full range ±2×(f/10)2 % *1 ±0.02×(f/10)2 % *1 4194A below 15 MHz Model 4274A ±2×(f/10)2 % *1 4275A Full range 4276A Full range ±0.02×(f/10)2 % *1 - - ±0.02% ±0.0002% 4277A*2 4278A*2 *1.
Operation ZERO OFFSET ADJUSTMENT ZERO OFFSET ADJUSTMENT The 16334A has inherent residual impedance and stray admittance that affects the accuracy of the measurements. To compensate for, or negate, these residuals in order to minimize measurement error, the instrument’s zero offset adjustment procedure should be performed. The zero offset adjustment procedure is described in the instrument’s manua1.
Operation OPERATION OPERATION Setup and measurement procedure is as follows: 1. Set the instrument’s cable LENGTH to 1m. 2. Connect the 16334A directly to the UNKNOUN terminals of the instrument. 3. Perform the ZERO offset adjustment described in the instrument’s operation manual. 4. Connect the DUT to the 16334A test fixture. WARNING DO NOT APPLY MORE THAN ±42V OF DC BIAS VOLTAGE TO THE UNKNOWN TERMINALS.
Operation OPERATION 16 Chapter 1
2 Service 17
Service Maintenance Maintenance An exploded view of the 16334A (for parts identification) is shown in Figure 2-1. Do not disassemble any further than shown. Maintenance consists principally of cleaning contacts and replacing worn or damaged parts. Take special care when cleaning contacts. To order parts, use the Agilent Technologies part numbers listed in Table 2-12-1.
Service Maintenance Figure 2-1 Parts Identification Chapter 2 19
Service Maintenance Table 2-1 Parts Identification Reference Agilent Part No. Qty.
Service Maintenance Table 2-1 Parts Identification Reference Agilent Part No. Qty. Description Note 29 16334-09003 2 SPACER 30 16334-00601 2 SHIELD 31 0890-0273 2x8.