Instruction manual

C
Copyright 2007 Agilent Technologies
Agilent 4263B LCR Meter Operation Manual
Manual Change
Agilent Part No. N/A
July 2008
Change 1
Change the Measurement Uncertainty value of Test Signal Frequency Accuracy Test (Page
9-38) to the following information.
Test Signal
Frequency
Test Limits Test Result
1
Measurement
Uncertainty
100 Hz ±0.010 Hz Hz ± 0.00031 Hz
120 Hz ±1.200 Hz Hz ± 0.00034 Hz
1 kHz ±0.00010 kHz kHz ± 0.0000029 kHz
10 kHz ±0.0010 kHz kHz ± 0.000019 kHz
20 kHz ±0.0010 kHz kHz ± 0.000020 kHz
100 kHz ±0.010 kHz kHz ± 0.00019 kHz
Change 2
Change the Measurement Uncertainty value of Test Signal Level Accuracy Test (Page 9-38)
to the following information.
Test Signal
Level
Test Signal
Frequency
Test Limits
Test Result
1
Measurement
Uncertainty
50 mV 100 Hz ±15.0 mV ± 0.90 mV
50 mV 20 kHz
2
±15.0 mV ± 0.078 mV
50 mV 100 kHz ±15.0 mV ± 0.18 mV
250 mV 100 Hz ±35.0 mV ± 2.50 mV
250 mV 20 kHz ±35.0 mV ± 0.35 mV
250 mV 100 kHz ±35.0 mV ± 1.88 mV
1 V 100 Hz ±0.11 V ± 0.00349 V
1 V 20 kHz ±0.11 V ± 0.00148 V
1 V 100 kHz ±0.11 V ± 0.00109 V