Instruction manual

C
Copyright 2009 Agilent Technologies
Measurement time:
Med
Test signal level:
500 mV
Status
PASS
Standard
Frequency [Hz]
Parameter Limits Result
1
Uncertainty
TAR
100 m
100
R ±0.62 m m ± 0.033 m
19.12
Note 1: Result = Measurement – Calibrated Value of Standard
Measurement:
Resistance
Cable length:
0 m
Measurement time:
Short
Test signal level:
500 mV
Status
PASS
Standard
Frequency [Hz]
Parameter Limits Result
1
Uncertainty
TAR
100 m
100
R ±1.10 m m ± 0.13 m
8.91
Note 1: Result = Measurement – Calibrated Value of Standard
Measurement:
DC Resistance
Cable length:
0 m
Measurement time:
Long
Test signal level:
1 V
Status
PASS
Standard Parameter Limits Result
1
Uncertainty TAR
100 m R ±0.97 m m
± 0.039
m
25.06
100 k R ±0.87 k m ± 0.013 k
67.56
Note 1: Result = Measurement – Calibrated Value of Standard