Instruction manual

C
Copyright 2007 Agilent Technologies
Change 19
Change the Measurement Uncertainty value of 4 m Impedance Measurement Accuracy Test
for 4 m DC Resistance Measurement Accuracy Test (Opt. 001.Only) (Page 9-45) to the
following information.
Standard
Test Limits
Test Result
1
Measurement
Uncertainty
100 mohm ±1.29 mohm mohm ± 0.030 mohm
100 kohm ±0.87 kohm kohm ± 0.013 kohm