Instruction manual

C
Copyright 2007 Agilent Technologies
Change 16
Change the Measurement Uncertainty value of 2 m Impedance Measurement Accuracy Test
for 2 m DC Resistance Measurement Accuracy Test (Opt. 001 Only) (Page 9-44) to the
following information.
Standard
Test Limits
Test Result
1
Measurement
Uncertainty
100 mohm ±1.13 mohm mohm ± 0.030 mohm
100 kohm ±0.87 kohm kohm ± 0.014 kohm
Change 17
Change the Measurement Uncertainty value of 4 m Impedance Measurement Accuracy Test
for 4 m Capacitance Measurement Accuracy Test (Page 9-45) to the following information.
Standard
Test Signal
Frequency
Parameter
Measured
Test Limits
Test Result
1
Measurement
Uncertainty
100 pF 1 kHz Cp ±0.19 pF pF ± 0.0078 pF
1 kHz D ±0.0019 ± 0.000065
1000 pF 100 Hz Cp ±0.002 nF pF ± 0.000089 nF
100 Hz D ±0.0020 ± 0.000062
1 μF 100 Hz Cp ±0.0018 μF μF ± 0.000077 μF
100 Hz D ±0.0018 ± 0.000047
1 μF 1 Hz Cp ±0.0011 μF μF ± 0.000071 μF
1 kHz D ±0.0011 ± 0.000033
Change 18
Change the Measurement Uncertainty value of 4 m Impedance Measurement Accuracy Test
for 4 m Resistance Measurement Accuracy Test (Page 9-45) to the following information.
Standard
Test Signal
Frequency
Parameter
Measured
Test Limits
Test Result
1
Measurement
Uncertainty
100 mohm 100 Hz R ±0.84 mohm mohm ± 0.030 mohm
100 mohm 1 kHz R ±0.96 mohm mohm ± 0.030 mohm