Change 11
Change the Measurement Uncertainty value of 0 m Impedance Measurement Accuracy Test
for 0 m DC Resistance Measurement Accuracy Test (Opt. 001 Only) (Page 9-42) to the
following information.
Standard
Test Limits
Test Result
1
Measurement
Uncertainty
100 mohm ±0.97 mohm mohm ± 0.039 mohm
100 kohm ±0.87 kohm kohm ± 0.013 kohm
Change 12
Change the Measurement Uncertainty value of 1 m Impedance Measurement Accuracy Test
for 1 m Capacitance Measurement Accuracy Test (Page 9-43) to the following information.
Standard
Test Signal
Frequency
Parameter
Measured
Test Limits
Test Result
1
Measurement
Uncertainty
100 pF 1 kHz Cp ±0.19 pF pF ± 0.0077 pF
D ±0.0019 ± 0.000063
100 pF 10 kHz Cp ±0.48 pF pF ± 0.012 pF
D ±0.0048 ± 0.00013
100 pF 20 kHz
2
Cp ±1.10 pF pF ± 0.024 pF
D ±0.0110 ± 0.000085
100 pF 100 kHz Cp ±1.37 pF pF ± 0.010 pF
D ±0.0137 ± 0.00016
1000 pF 100 Hz Cp ±0.002 nF pF ± 0.000091 nF
D ±0.0020 ± 0.000063
1 μF 100 Hz Cp ±0.0018 μF μF ± 0.000075 μF
D ±0.0018 ± 0.000047
1 μF 1 kHz Cp ±0.0011 μF μF ± 0.000069 μF
D ±0.0011 ± 0.000033
1 μF 10 kHz Cp ±0.0026 μF μF ± 0.00014 μF
D ±0.0026 ± 0.000061
1 μF 100 kHz Cp ±0.0207 μF μF ± 0.00016 μF
D ±0.0207 ± 0.00029
Change 13
Change the Measurement Uncertainty value of 1 m Impedance Measurement Accuracy Test
for 1 m Resistance Measurement Accuracy Test (Page 9-43) to the following information.
Standard
Test Signal
Frequency
Parameter
Measured
Test Limits
Test Result
1
Measurement
Uncertainty
100 mohm 100 Hz R ±0.60 mohm mohm ± 0.030 mohm
100 mohm 1 kHz R ±0.60 mohm mohm ± 0.031 mohm