4 Reconnect the power cord. Turn on the instrument by setting the
power switch to the On position.
To degauss the display
If the logic analyzer has been subjected to strong magnetic fields, the
CRT might become magnetized and display data might become
distorted. To correct this condition, degauss the CRT with a
conventional external television type degaussing coil.
To clean the logic analyzer
With the instrument turned off and unplugged, use mild soap and
water to clean the front and cabinet of the logic analyzer. Harsh soap
might damage the water-base paint.
To test the logic analyzer
If you require a test to verify the specifications, the Agilent
Technologies 1660 Series Logic Analyzer Service Manual is
required(Agilent Technologies 1660 Series Logic Analyzer Service
Manual for the 1664A). Start at the beginning of chapter 3, "Testing
Performance."
If you require a test to initially accept the operation, perform the
self-tests described in Troubleshooting in this chapter.
If the logic analyzer does not operate correctly, go to the flow charts
provided in Troubleshooting in this chapter.