Technical data

Advanced Triggering, Search & Navigation, and Segmented Acquisition Labs 3
4000 X-Series Oscilloscopes Advanced Training Guide 71
Your scope’s display should now look similar to Figure 44 while using the
scope’s default trigger condition (rising edge of channel- 1). Let’s now set
up the scope to trigger on one of the wider digital pulses on channel- 2. To
do this we will qualify triggering on the channel-1 pulse, then delay
arming of the trigger past the 5 MHz analog burst, and then finally trigger
on the Nth occurrence of one of the wider digital pulses.
12 Press the [Trigger] front panel key; then select Edge then Edge.
13 Double- tap the Delay softkey; then enter 2.0 µs.
14 Double- tap the Nth Edge B softkey and enter 3.
Figure 44 Capturing a complex burst signal while triggering on a synchronization signal.