Serial Bus Decoding & Triggering, Search & Navigation, and Segmented Acquisition Labs 4
4000 X-Series Oscilloscopes Advanced Training Guide 147
12 Set the scope timebase to 200.0 µs/div.
13 Press the [Search] feature in the Horizontal area of the front panel.
14 Tap the Search Edge softkey and change the mode to Serial 1:
MIL-STD-1553.
15 Open the Lister by tapping the double down arrows to the upper right of
the waveform display.
White arrows should appear on the top of the waveform, below the Lister
window. These represent the locations of all hits from the search
parameters, namely all command and segment status word starts. There
should be 10- 11 on the display. They are also marked in the Lister
window with orange tabs on the left column.
16 Tap Search for C/S Start and change it to Sync Error.
17 Press [Run/Stop] to halt all acquisitions.
18 Use the and front panel navigation keys to automatically scroll
through all search results. The lower screen will automatically zoom
into the sync error in the search list, and point to it in the lister
window. See Figure 101.
Figure 100 Triggering on a unique sync error.