Technical data

Serial Bus Decoding & Triggering, Search & Navigation, and Segmented Acquisition Labs 4
4000 X-Series Oscilloscopes Advanced Training Guide 129
Lister Menu on the top right of the signal list, and tap the Zoom to Selection
softkey. Let's now perform an automatic search to find every occurrence of
a particular Frame ID. We will then automatically navigate to each of
these occurrences.
20 Push the horizontal position/delay knob to re- position the trigger point
back to center-screen.
21 Set the scope's timebase to 100.00 µs/div.
22 Press [Run/Stop] to restart repetitive acquisitions.
23 Press the [Search] front panel key.
24 Tap the Search Edge softkey; then select Serial 1: FlexRay.
25 Tap the Search for softkey; then select Frame ID as our search criteria.
26 Tap the Frame ID softkey; enter 1 as our search criteria.
The white triangles near the top of the waveform area mark the time
location of each "found" occurrence of our search operation as shown in
Figure 87. These frames are also marked in orange in the first column of
the lister table. Your scope should have found and marked approximately
20 occurrences of this search operation based on a total acquisition time
of 1 ms. Note that the number of "found" events is indicated near the
bottom of the display. If you want to automatically navigate to each frame
with ID of 1, press the and front panel navigation keys (near the
timebase controls).