4 Troubleshooting
To perform hardware self test
Pressing [Utility] > Service > Hardware Self Test performs a series
of internal procedures to verify that the oscilloscope is
operating properly.
It is recommended you run Hardware Self Test:
After experiencing abnormal operation.
For additional information to better describe an
oscilloscope failure.
To verify proper operation after the oscilloscope has been
repaired.
Successfully passing Hardware Self Test does not guarantee
100% of the oscilloscope's functionality. Hardware Self Test
is designed to provide an 80% confidence level that the
oscilloscope is operating properly.
To perform front panel self test
Pressing [Utility] > Service > Front Panel Self Test lets you test the
front panel keys and knobs as well as the oscilloscope
display.
Follow the on- screen instructions.
Failures in the front panel self test indicate problems with
the keyboard, keypad, or display.