Technical data

Troubleshooting 4
4000 X-Series Oscilloscopes Service Guide 73
If there is no trace display
Check that the Intensity (on the front panel) is adjusted
correctly.
Recall the default setup by pressing [Default Setup]. This
will ensure that the trigger mode is Auto.
Check that the probe clips are securely connected to
points in the circuit under test, and that the ground is
connected.
Check that the circuit under test is powered on.
Press the [AutoScale] key.
Obtain service from Agilent Technologies, if necessary.
If the trace display is unusual or unexpected
Check that the Horizontal time/division setting is correct
for the expected frequency range of the input signals.
The sampling speed of the oscilloscope depends on the
time/division setting. It may be that when time/division is
set to slower speeds, the oscilloscope is sampling too
slowly to capture all of the transitions on the waveform.
Use peak detect mode.
Check that all oscilloscope probes are connected to the
correct signals in the circuit under test.
Ensure that the probe’s ground lead is securely connected
to a ground point in the circuit under test. For
high- speed measurements, each probe’s individual ground
lead should also be connected to a ground point closest to
the signal point in the circuit under test.
Check that the trigger setup is correct.
A correct trigger setup is the most important factor in
helping you capture the data you desire. See the User’s
Guide for information about triggering.