Technical data
Testing Performance 2
4000 X-Series Oscilloscopes Service Guide 63
External Trigger Sensitivity
Input Range: 8 V Generator Setting Test Limits Ext Trig In
200 MHz 350 mV ________
100 MHz 200 mV ________
Input Range: 1.6 V Generator Setting Test Limits Ext Trig In
200 MHz 70 mV ________
100 MHz 40 mV ________
* Where applicable