Technical data

Testing Performance 2
4000 X-Series Oscilloscopes Service Guide 57
Test External Trigger Sensitivity (all models)
This test applies to all models.
Verify the external trigger sensitivity at these settings:
1 Connect the equipment (see Figure 9).
a Use the N cable to connect the signal generator to the
power splitter input.
b Connect one output of the power splitter to the
Aux Trig input through a 50Ω feedthrough termination.
c Connect the power sensor to the other output of the
power splitter.
Table 17 External Trigger Sensitivity Test Settings
Input Range 100 MHz 200 Mhz
1.6 V 40 mV
pp
70 mV
pp
8V 200mV
pp
350 mV
pp