Technical data
Testing Performance 2
4000 X-Series Oscilloscopes Service Guide 53
To verify trigger sensitivity
This test verifies the trigger sensitivity. In this test, you will
apply a sine wave to the oscilloscope at the upper
bandwidth limit. You will then decrease the amplitude of the
signal to the specified levels, and check to see if the
oscilloscope is still triggered.
Table 14 Internal Trigger SensitivityTest Limits
BW Models V/div Frequency Sensitivity
200 MHz,
350 MHz,
500 MHz, and
1GHz
< 10 mV/div All greater of 1 div or 5 mV
pp
>= 10 mV/div All 0.6 div
1.5 GHz < 10 mV/div DC to 1 GHz greater of 1 div or 5 mV
pp
1 GHz to 1.5 GHz greater of 1.5 div or 5 mV
pp
>= 10 mV/div DC to 1 GHz 0.6 div
1 GHz to 1.5 GHz 1.0 div
Table 15 External Trigger Sensitivity Test Limits, All Models
Input Range Frequency Sensitivity
1.6 V DC to 100 MHz 40 mV
pp
100 MHz to 200 MHz 70 mV
pp
8V DC to 100MHz 200mV
pp
100 MHz to 200 MHz 350 mV
pp