Technical data
Testing Performance 2
4000 X-Series Oscilloscopes Service Guide 41
c Press the [Acquire] key.
d Then press the Acq Mode softkey and select Averaging.
e Then press the #Avgs softkey and set it to 64.
Wait a few seconds for the measurement to settle.
3 Press the [Cursors] key, set the Mode softkey to Normal,
then press the XY softkey and select Y. Press the Y1
softkey, then use the Entry knob (labeled on the
front panel) to set the Y1 cursor on the baseline of the
signal.
Table 10 Settings Used to Verify Dual Cursor Accuracy
Volts/Div
Setting
Oscilloscope
Calibrator Setting
Test Limits
5 V/Div 35 V 34.032 V to 35.968 V
2 V/Div 14 V 13.6128 V to 14.3872 V
1V/Div 7V 6.8064V to 7.1936V
500 mV/Div 3.5 V 3.4032 V to 3.5968 V
200 mV/Div 1.4 V 1.36128 V to 1.43872 V
100 mV/Div 700 mV 680.64 mV to 719.36 mV
50 mV/Div 350 mV 340.32 mV to 359.68 mV
20 mV/Div 140 mV 136.128 mV to 143.87 mV
10 mV/Div 70 mV 68.064 mV to 71.94 mV
5mV/Div
1
35 mV 34.032 mV to 35.97 mV
2mV/Div
1, 2
14 mV 13.226 mV to 14.77 mV
1mV/Div
1, 2
7 mV 6.226 mV to 7.77 mV
1
A blocking capacitor is required at this range to reduce noise. See “Use a
Blocking Capacitor to Reduce Noise” on page 44.
2
Full scale is defined as 32 mV on the 2 mV/div range and the 1 mV/div range.
Full scale on all other ranges is defined as 8 divisions times the V/div setting.