Technical data

24 4000 X-Series Oscilloscopes Service Guide
2 Testing Performance
Conventions
The following conventions will be used when referring to
oscilloscope models throughout this chapter.
Adapter Type N (m) to BNC (m) Agilent 1250-0082 or
Pomona 3288 with
Pomona 3533
Blocking capacitor and shorting cap Note: if a BNC blocking capacitor is not available use an
SMA blocking capacitor.
Agilent 11742A +
Pomona 4288 +
Pomona 5088
Adapter (qty 3) N(m) to BNC(f) Agilent 1250-0780
50 Ohm Feedthrough Termination 50Ω BNC (f) to BNC (m) Agilent 0960-0301
Table 1 List of test equipment (continued)
Equipment Critical Specifications
Recommended Model/
Part Number
* Required only for testing digital channels of oscilloscopes that have the MSO option.
Most parts and equipment are available at www.agilent.com. See respective manufacturer’s websites for their equipment.
These parts available at www.parts.agilent.com at the time this manual was published.
Table 2 Conventions
Models Referred to as:
MSO-X /DSO-X 4022A, MSO-X /DSO-X 4024A 200 MHz Models
MSO-X /DSO-X 4032A, MSO-X /DSO-X 4034A 350 MHz Models
MSO-X /DSO-X 4052A, MSO-X /DSO-X 4054A 500 MHz Models
MSO-X /DSO-X 4104A 1 GHz Models
MSO-X /DSO-X 4154A 1.5 GHz Models