Technical data
Testing Performance 2
4000 X-Series Oscilloscopes Service Guide 23
List of Test Equipment
Below is a list of test equipment and accessories required to
perform the performance test verification procedures.
Table 1 List of test equipment
Equipment Critical Specifications
Recommended Model/
Part Number
Test connector, 8-by-2* See page 25 for instructions on building test connector. n/a
Digital Multimeter 0.1 mV resolution, 0.005% accuracy Agilent 34401A
Power Splitter Outputs differ by 0.15 dB Agilent 11667B
Oscilloscope Calibrator DC offset voltage of -5.5 V to 35.5 V, 0.1 V resolution
25 MHz—500 MHz sine wave, 5 ppm
Fluke 5820A
Signal Generator 100 kHz to 1.5 GHz sine waves Agilent N5181A with
3GHz option
Power Meter 1.5 GHz
±3% accuracy Agilent N1914A
Power Sensor 1.5 GHz
±3% accuracy Agilent E9304A or
N8482A
BNC banana cable BNC (m) to dual banana Pomona 2BC-BNC-36 or
Agilent 11001-66001
BNC cable (qty 3) BNC - BNC, 48” length Agilent 10503A
†
Cable Type N (m) 609.6 mm (24 in.) Agilent 11500B
Probe cable* No substitute Agilent N6450-60001
(16-channel) or Agilent
N6459-60001
(8-channel)
†
Adapter BNC(f) to banana(m) Agilent 1251-2277
†
Adapter BNC Tee (m) (f) (f) Agilent 1250-0781
†
or
Pomona 3285
* Required only for testing digital channels of oscilloscopes that have the MSO option.
Most parts and equipment are available at www.agilent.com. See respective manufacturer’s websites for their equipment.
†
These parts available at www.parts.agilent.com at the time this manual was published.