Converts a binary setup state file from the HP 3563A (or HP 3562A) to an ASCII form (written to
destination file). Refer to the HP 3563A or HP 3562A programming manual for descriptions of the
individual fields in the setup state file. HP 3563A/62A setup state files on a LIF file system start with the
letters ‘’SE.’’
Syntax SETUP63 <sfile> [dfile] [/U] [/O] [/L:<flag>] [/E:<flag>]
Options
<sfile> Input HP 3563A/62A setup state file name.
[dfile] Optional output ASCII file name. If no file specified, output to display.
/U Show help (usage information) for this program.
/O Overwrite [dfile] if it exists.
/L:<flag> If <flag> is non-zero, put labels on each line. The name of each field will be the
first 25 characters of each line (approximately as they appear in the HP 3563A or
HP 3562A Programming Manual). The default is on.
/E:<flag> If <flag> is non-zero, expand enumerated types to ASCII names (approximately as
they appear in the HP 3563A or HP 3562A Programming Manual). The default is
on.
Sharing Data Between Analyzers