Technical data

Service Guide 4
Agilent 35670A Supplement 27
To troubleshoot memory or battery failures
Replace the last five bulleted items:
If the date is 01- 01-01, the battery- backed- up memory is
functioning correctly. Enter the current date. Go to page
4- 13, “To perform self tests,” to continue troubleshooting.
Set power switch to off (O)
Check that the voltage at TP200 is 2.6Vdc to 3.1Vdc.
If the voltage is correct, the memory is probably faulty.
If the voltage is incorrect, replace the battery BT1.
Page in original guide 4-67,68
Figure 12 Battery location
Battery
TP200