To troubleshoot display failures
Replace entire section
Use this test to isolate display failures to the A101 Display
assembly, A102 assembly, or the A17 CPU assembly
Step 1. Check the power inverter assembly.
Set the power switch to off (O).
Remove the cable connected to P13 on the A17 assembly.
Set the power switch to on (|)
Measure voltage at P13 pin 1 on A17 assembly.
If voltage is not +5Vdc the A17 board may be faulty.
Set the power switch to off (O).
Reconnect the cable to P13.
Set the power switch to on (|).
Measure voltage at A17 P13 pin 1.
If voltage is not +5Vdc the power inverter assembly may be
faulty.
If display backlight lamps do not light up, one or both of the
lamps may be faulty.
If lamps are replaced and still do not light up, the power
inverter may be faulty.
Step 2. Check the CPU signals on the Display assembly.
Set the power switch to on (|).
Using a logic probe check that the following TTL signals are
toggling:
If the signals are incorrect the A17 board may be faulty.
Page in original guide 4-23
Test Location Signal Na m e In/O u t
P14-31 VSYNCLCD A17 Out
P14-33 HSYNCLCD A17 Out
P14-24 CLK_LCD A17 Out