Technical data
20 Agilent 35670A Supplement
4 Service Guide
To troubleshoot power-up failures
Replace the first five rows of the table:
Page in original guide 4-15
Failing Power-up Message Probable Faulty Assembly or Next Test
LEDs
MC68030 Processor
MC68882 Coprocessor
Bootrom
Display
A17 CPU
Main RAM
ProgramROM
A17 CPU
DSP A17 CPU
Fast bus A17 CPU
MFP A17 CPU