Programming instructions

70 Verification Tests
Correction Replace the DIO module with a new one if it is failed.
Agilent 44472A Verification Tests
The VHF Switch module consists of two 4-to-1 MUXs with BNC
connections. Only the center conductor of each channel is switched. The
outer conductors (shield) of all channels in a particular MUX are connected
together and unswitched.
Tests Procedure To test a channel:
1. Remove any external wiring from the rear of the 44472A. Reset the
44472A and press Mon;
2. Set the Multimeter to measure resistance, “Zero” the Multimeter;
3. Create a scan list including all the channels of 44472A;
4. MUX #0 TESTS: Connect a Multimeter between the center
conductor of the common connector (common #0) and the center
conductor of the channel under test (channel 00 to start). The
Multimeter should indicate that the channel is open (very high
resistance or an overload indication);
5. Press Step, the Multimeter should display a resistance reading of less
than 1;
6. Repeat Steps 4 and 5 for channels 01, 02 and 03;
7. MUX #1 TESTS: Reset the 44472A and connect the Multimeter
between the center conductor of the common connector (common #1)
and the center conductor of the channel under test (channel 10 to
start). The Multimeter should indicate that the channel is open (very
high resistance or an overload indication);
8. Press Step until the channel under test appears in the Agilent
3499A/B/C’s display. The Multimeter should display a resistance
reading of less than 1;
9. Repeat Steps 7 and 8 for channels 11, 12 and 13.
Correction a. If during Steps 5 or 8, one or more channels does not close as described,
a problem exists in either the mainframe Controller board or the 44472A.
You can use a functional 44472A to isolate the problem.
b. If during Steps 5 or 8, the channels closed but one or more of the
resistances was greater than 1, the relays in the corresponding channel(s)
are failing.
Note Replace the 44472A with a new one if it is failing.
Agilent 44476A/B Verification Tests
The Agilent 44476A has three independent SPDT 50 coaxial switches
(Figure 4-5). The Agilent 44476B allows you to mount any two 3331XX