42 Front-Panel Operation
3. Press Enter, turn the knob to select one mode (i.e. WIRE1);
4. Press Enter to return to “CONFIG MUX”, press Mode again to exit
the configuration, the CONFIG annunciator turns off accordingly.
DIO Module
Configuration
A digital I/O module can be configured, which include the flow control
mode, the control line polarity, the flag line polarity, and the I/O direction
line polarity.
To configure a digital I/O module:
1. Select a slot in which a digital I/O is installed;
2. Press Mode, the CONFIG annunciator lights up and “CONFIG DIO”
is displayed, then press Enter;
3. Turn the knob to select the flow control mode (i.e. MODE 2),
then press Enter;
4. Turn the knob to select the control line polarity (i.e. CONT POL
POS), then press Enter;
5. Turn the knob to select the flag line polarity (i.e. FLAG POL POS),
then press Enter;
6. Turn the knob to select the I/O line polarity (i.e. I/O POL POS),
then press Enter;
WIRE1 1
N2263A 4
CONFIG DIO 4
MODE 2 4
CONT POL POS 4
FLAG POL POS 4
I/O POL POS
4