Programming instructions
10 Specifications
Matrix Modules The matrix modules include: Agilent N2262A 4 x 8 2-wire matrix, Agilent
44473A 4 x 4 2-wire matrix and the 4 x 4 2-wire matrix on an Agilent
N2265A. The specifications are listed in the table below.
ITEMS SPECIFICATIONS
Agilent N2262A
Agilent N2265A
a
(4 x 4 Matrix)
Agilent 44473A
INPUT CHARACTERISTICS
Total Channels: 4 x 8 4 x 4 4 x 4
Maximum Voltage Terminal-Terminal or
Terminal-Chassis:
200 V, dc or ac rms 200 V, dc or ac rms 250 V, dc or ac rms
Maximum Current
Per Channel:
Per Module:
1 A, dc or ac rms
4 A, dc or ac rms
1 A, dc or ac rms
4 A, dc or ac rms
2 A, dc or ac rms
8 A, dc or ac rms
Maximum Power
Per Channel:
Per Module:
60 W dc; 62.5 VA ac
240 W dc; 250 VA ac
60 W dc; 62.5 VA ac
240 W dc; 250 VA ac
60 W dc; 500 VA ac
240 W dc; 2000 VA ac
Maximum Overvoltage Transients
N/A N/A 1400 V
pk
Thermal Offset: < 3 µV differential < 3 µV differential < 3 µV differential
Initial Closed Channel Resistance: < 1 Ω < 1 Ω < 1 Ω
Relay Life
Mechanical: 10
8
(at 36000
operations/hour)
Electrical: 5 x 10
5
(1A load)
Mechanical: 10
8
(at 36000 operations/hour)
Electrical: 5 x 10
5
(1A load)
10
8
operations/chan (dry load
of < 300 mA & < 10 V)
10
5
operations/chan
(maximum rated load)
Maximum Scan Rate: 80 Chans/sec 80 Chans/sec 43 Chans/sec
DC ISOLATION (with terminal block)
Open Channel,
Channel-Channel
(with 1 channel closed)
< (40
o
C, 50% RH): > 10
10
Ω
< (40
o
C, 80% RH): > 10
9
Ω
< (40
o
C, 50% RH): > 10
10
Ω
< (40
o
C, 80% RH): > 10
9
Ω
< (40
o
C, 60% RH): > 10
11
Ω
< (40
o
C, 95% RH): > 10
9
Ω
HI-LO
(with 1 channel closed)
< (40
o
C, 50% RH): > 10
10
Ω
< (40
o
C, 80% RH): > 10
9
Ω
< (40
o
C, 50% RH): > 10
10
Ω
< (40
o
C, 80% RH): > 10
9
Ω
< (40
o
C, 60% RH): > 10
10
Ω
< (40
o
C, 95% RH): > 10
8
Ω
Channel-Chassis
(with 1 channel closed)
< (40
o
C, 50% RH): > 10
10
Ω
< (40
o
C, 80% RH): > 10
9
Ω
< (40
o
C, 50% RH): > 10
10
Ω
< (40
o
C, 80% RH): > 10
9
Ω
< (40
o
C, 60% RH): > 10
10
Ω
< (40
o
C, 95% RH): > 5x10
8
Ω
AC ISOLATION / PERFORMANCE
b
(without terminal block)
Capacitance
(with 1 channel closed)
Open Channel,
Channel-Channel:
HI-LO:
Channel-Chassis:
< 7 pF
< 30 pF
< 50 pF
< 7 pF
< 25 pF
< 40 pF
< 5 pF
< 40 pF
< 70 pF
Insertion Loss
(with 50
Ω termination)
100 kHz:
1 MHz:
10 MHz:
< 0.10 dB
< 0.20 dB
< 0.60 dB
< 0.10 dB
< 0.20 dB
< 0.60 dB
< 0.30 dB
< 0.35 dB
< 0.90 dB
Crosstalk
(with 50Ω termination)
100 kHz:
1 MHz:
10 MHz:
< -73 dB
< -53 dB
< -28 dB
< -76 dB
< -56 dB
< -33 dB
< -76 dB
< -56 dB
< -36 dB
a. Agilent N2265A combines a 4 x 4 matrix and a 16-bit digital I/O into one module. It can be operated as two separate modules.