Programming instructions
128 Error Messages
In 3488A Mode
Error Conditions
In 3488A mode, you can also query the error queue when the ERROR annuciator is on. The returned decimal
value is equal to the sum of the values of the possible error conditions, as defined in Table 6-20.
Note
A string “+00000” read back (with the command ERROR) indicates that the error queue is
empty. In this case, a string “0000” displays on the front panel of the instrument.
Self Test Errors
The errors listed in Table 6-21 indicate failures that may occur during a self-test (in 3488A mode).
Note
A string “+0” read back (with the command TEST) indicates that all the tests have passed. In
this case, a string “PASSED” displays on the front panel of the instrument.
Table 6-20. Error Conditions in 3488A Mode
Weighted Value Error Condition
1 Syntax Error
2 Execution Error which include:
a. Parameter out of range;
b. Module type mismatch;
c. Attempt to access a nonexistent stored state
or scan list.
4 Hardware Trigger too fast
8 Logical Failure
16 Power Supply Failure
Table 6-21. Self -Test Errors
Error Number Description
+1 ROM test failed.
+2 GPIB test failed.
+3 RS-232 test failed.
+4 Front-panel test failed.