118 Plug-in Modules Wiring Information
Agilent N2320A
Wiring Sequence
Figure 6-26. Agilent N2320A Connector Wiring Sequence
Step 3.
Step 2.
Step 1.
Release the four screws and uncover the upper metal shield
Rear view Front view
Metal strain relief clamp
Upper metal shield
Insert wired Socket into the Receptacle
Replace the metal shield and tighten relevant screws
The strain relief clamp can be used to tighten wires with additional insulation around them. Additional insulation should
be used around the wires to prevent possible damage caused by the strain relief clamp.
Stripped wire (20-24 AWG) Socket
A. Prepare the Wire and the Socket.
B. Insert the Wire into the Socket.
C. Crimp the Wire with the Socket by using
an AMP Hand Crimping Tool (90067-5)
or an equivalent tool.
Plastic isolation tube