User`s guide
24
Additional Low Frequency Error for ACV, ACI (% of reading)
Frequency
AC Filter Slow AC Filter Medium AC Filter Fast
10Hz-20Hz 0 0.74 –
20Hz-40Hz 0 0.22 –
40Hz-100Hz 0 0.06 0.73
100 Hz - 200 Hz 0 0.01 0.22
200 Hz - 1 kHz 0 0 0.18
>1kHz 0 0 0
Additional Error for Frequency, Period (% of reading)
Aperture (Digits)
Frequency
1 second 0.1 seconds 0.01 seconds
( 6
1
/
2
digits) (5
1
/
2
digits) (4
1
/
2
digits)
3 Hz- 5 Hz 0 0.12 0.12
5Hz-10Hz 0 0.17 0.17
10 Hz - 40 Hz 0 0.2 0.2
40 Hz - 100 Hz 0 0.06 0.21
100 Hz - 300 Hz 0 0.03 0.21
300Hz-1kHz 0 0.01 0.07
>1kHz 0 0 0.02
Typical system speeds (Measurementsmadeona3.2GHzPCrunningVB6inWindowsXPPro)
Single Channel Reading times in msec Direct measurements – direct to I/O Direct measurement
(includes switch, measure time and I/O time) to memory(GPIB)
Single channel
[1] [2]
GPIB USB2.0 LAN Measurementinto
msec msec (w/ VXI 11) msec memory msec
Singlechannel,DCV 2.83 3.14 4.57 1.90
Singlechannel,ACV 5.00 5.35 5.75 4.00
Single channel, ohms 2.91 3.14 4.65 1.90
Singlechannelwhilechanging 9.52 10.64 11.76 8.40
scale (eg MEAS DCV 10 / MEAS DCV 1)
Single channel while changing 128 120 120 120
function (eg. MEAS ACV / MEAS DCV)
Command execution time
[3]
34925A OpenorClose 0.7 0.9 1.6
Read? 2.9 3.3 4.7
Close/Read/Open 4.8 5.3 6.5
Init/*WAI 1.9 2.1 3
Close/Init/Open 3.7 4.1 4.7
34923A Open or Close 0.9 1.2 1.8
Read? 2.9 3.3 4.7
Close/Read/Open 5.3 5.8 6.5
Init/*WAI 1.9 2.1 3
Close/Init/Open 4.2 4.7 5.2
34921A OpenorClose 4.7 5 5.3
Read? 2.9 3.3 4.7
Close/Read/Open 14 15 15
Init/*WAI 1.9 2.1 3
Close/Init/Open 12.4 14 14
34934A Close 1.8 2.2 3.1
Read? 2.8 3.1 4.8
Open 1.5 1.8 3.2
[1] ReadingsweremadewithminimumNPLC,delay0,displayoff,autozerooff
[2] Alltimesincludetheissueof“READ?”andtheretrievalofdata
[3] CLOSEorOPENbustransfertimesallowedtooverlappreviouscommand;commandparsetimesoverlapcurrentactivityuntilIOlatencydominant
Temperature measurement accuracy
Temperature 1-year accuracy Type Best range
[1]
Extended range
[1]
Temperature
coefficient/°C
Thermocouple B 1100°Cto1820°C 1.2°C 400°Cto1100°C 1.8°C 0.03°C
(34921A only, E -150°C to 1000°C 1.0°C -200°C to -150°C 1.5°C 0.03°C
includes cold J -150°C to 1200°C 1.0°C -210°C to -150°C 1.2°C 0.03°C
junction accuracy K -100°C to 1200°C 1.0°C -200°C to -100°C 1.5°C 0.03°C
onterminalblock) N -100°Cto1300°C 1.0°C -200°Cto-100°C 1.5°C 0.03°C
R 300°Cto1760°C 1.2°C -50°Cto300°C 1.8°C 0.03°C
S 400°Cto1760°C 1.2°C -50°Cto400°C 1.8°C 0.03°C
T -100°C to 400°C 1.0°C -200°C to -100°C 1.5°C 0.03°C
RTD R
o
from -200°C to 600°C 0.06°C 0.003°C
49Ωto2.1KΩ
Thermistor 2.2 k, 5 k, 10 k -80°C to 150°C 0.08°C 0.002°C
[1] For total measurement accuracy, add temperature probe error; for <1
°
C accuracy, an external fixed reference is required