User`s guide
14
34980A general-purpose
switch modules
The 34980A general-purpose switches can
be used to route signals or to control other
system devices. These switches are ideal
for device actuation and switching loads or
power supplies.
Choose from the following features:
• FormCchannelsupto1A,50W
• FormAchannelsupto5A,150W
• Armaturelatchingrelays
• Simultaneouschannelswitching
• Temperaturesensortodetect
overheating conditions
• Connectionsviastandard50or78-pin
Dsub cables or detachable terminal block
The34937Aisthemostversatilegeneral-
purpose switch with 28 Form C channels that
can switch up to 1 A of current. In addition,
this module has four Form A channels that
can switch up to 5 A of current. For power
switching applications, the 34938A has 20
5-amp channels in a Form A topology. Each
Form A general-purpose switch can handle
up to 150 W, enough for many power line-
switching applications. For high density
applications the 34939A offers 64 Form A
channels for switching up to 1A and carry
currents up to 2A.
The general purpose switches contain
latching armature relays where multiple
channels can be closed at the same time.
Additionally, for switching reactive loads,
the optional terminal blocks have pads for
snubbing circuits.
The built-in relay counter helps predict
when relays need to be replaced.
Table 5. GP actuator selection table—specifications and characteristics
34937A 34938A 34939A
Channels/configurations 28 Form C 20 Form A 64 Form A
4 Form A
Switch type Armature, latching Armature, latching Armature, latching
Input characteristics (per channel)
Maxvolts(DC,ACRMS)
[1]
FormC–300V 30VDC/250VAC +/-100Vpeak
FormA–30VDC/250VAC
Maxcurrent(DC,ACRMS) FormC–1A(2Acarry) 5Aswitch 1Aswitch
FormA–5Aswitch (8Acarry) (2Acarry)
(8 A carry)
Power (W, VA)
[2]
FormC–60W 60W
FormA–150W 150W
Volt-Hertz limit 10
8
10
8
10
8
General specifications
Offset voltage 3 uV 3 uV 3 uV
Initialclosedchannelres FormC–125mΩ <125mΩ
FormA–50mΩ <60mΩ
DCIsolation(ch-ch,ch-earth) >10GΩ >10GΩ 10GΩ
AC characteristics
Bandwidthatterminalblock[3] 10MHz 1MHz 10MHz
Channel Isolation at terminal block [3]
100kHz 55dB 60dB 45dB
1MHz 35dB 40dB 25dB
10MHz 15dB 5dB
Capacitance at terminal block
CH–CH FormC12pF/FormA10pF 65pF 20pF
CH–earth FormC21pF/FormA18pF 105pF 70pF
General characteristics
Relaylifenoload/rated FormC–100M/100k >100M/100k
FormA–50M/30k 50M/30k
Open/closetime FormC–4ms/4ms 10ms/10ms 4ms/4ms
FormA–10ms/10ms
Initial/resetrelaystate FormC–maintainstate userconfigurable maintain
FormA–userconfigurable
Analogbusbackplaneconnection No No No
[1] DCorACRMSvoltage,channel-to-channelorchannel-to-earth
[2] Limitedto6Wofchannelresistancepowerlosspermodule
[3] 50Ωsource,50Ωload,differentialmeasurementsverified(S21)
Channel 001
(1A form C relay)
NC
NO
COM
Channel 028
(1A form C relay)
NC
NO
COM
Channel 029
(5A form A relay)
NO
COM
Channel 032
(5A form A relay)
NO
COM
Figure 9. 34937A 32-channel Form A/
Form C switch