Technical data

Troubleshooting and Diagnostics 4
34980A Service Guide 79
Power Supply
The main power supply provides +12 Volts. All other power supplies are
derived from this supply. To test the main power supply:
1 Disassemble the instrument as described beginning on page 115.
2 Use a DVM to check that the power supply output is +12 V ± 0.6 V.
Test the power supply output at the red leads on the power supply
connector to the communications board.
a The +12 Volt supply is always on when the ac power is applied to
the instrument. Test for loading errors by unplugging the power
supply connector from the communications board.
3 Use an oscilloscope to verify there is no ac oscillation on the +12 V
power supply.
Replace the power supply assembly if +12 V is not present.
Product Firmware Updates
As new product features and enhancements become available, you can
easily update your mainframe and plug- in module firmware to ensure
optimum compatibility. The latest firmware updates are available from the
Agilent 34980A product page at www.agilent.com/find/34980AUpdates.
Front Panel Operation: Utility > FIRMWARE > UPDATE
Once you have downloaded the latest mainframe firmware (see above),
use the knob to scroll through the installed modules that require a
firmware update. To exit the menu without installing the updates,
select CANCEL.
WARNING
Exposed Mains. When the instrument cover is removed to test
the power supply, the ac mains are exposed.