3 Calibration Procedures
Optional AC Performance Verification Tests
These tests are not intended to be performed with every calibration. They
are provided as an aid for verifying additional instrument specifications.
There are no adjustments for these tests; they are provided for
performance verification only.
Configuration: AC Volts
CONFigure[:VOLTage]:AC
LF 3 HZ:SLOW
[SENSe:]VOLTage:AC:BANDwidth 3
1
Make sure you have read “DMM Test Considerations" on page 53.
2 Select the AC Volts function and the 3 Hz filter.
3 Select each range in the order shown below. Provide the input voltage
and frequency indicated. Compare measurement results to the
appropriate test limits shown in the table below. (Be certain to allow
for appropriate source settling.)
Input Error from Nominal
Voltage Frequenc
y
Range 24 hour 90 day 1 year
1 V 20 Hz 1 V ± 700 µV ± 900 µV ± 1 mV
1 V 20 kHz 1 V ± 700 µV ± 900 µV ± 1 mV
1 V 100 kHz 1 V ± 6.3 mV ± 6.8 mV ± 6.8 mV
1 V 300 kHz 1 V ± 45 mV ± 45 mV ± 45 mV
10 V 1 kHz 10 V ± 7 mV ± 9 mV ± 10 mV
1 V 1 kHz 10 V ± 3.4 mV ± 4.5 mV ± 4.6 mV
100 mV 1 kHz 10 V ± 13 mV ± 14 mV ± 14 mV