Technical data

58 34980A Service Guide
3 Calibration Procedures
AC CURRENT Gain Verification Test
Configuration: AC Current
CONFigure:CURRent:AC
LF 3 HZ:SLOW
[SENSe:]CURRent:AC:BANDwidth 3
1
Make sure you have read “DMM Test Considerations" on page 53.
2 Set the AC CURRENT function and the 3 Hz input filter. With the slow
filter selected, each measurement takes 7 seconds to complete.
3 Select each range in the order shown below. Provide the input current
and frequency indicated. Compare measurement results to the
appropriate test limits shown in the table. (Be certain to allow for
appropriate source settling.)
[1] Verify only, no adjustment.
Q: Quick performance verification test points.
NOTE
The 50 kHz ac voltage test points may fail performance verification if
the DMM internal shields have been removed and reinstalled. See “Gain
Adjustment" on page 61 for further information on how to recalibrate
the ac voltage function.
Input Quick
Check
Error from Nominal
Current Frequency Range 24 hour 90 day 1 year
10 mA
[1]
1 kHz 10 mA ± 14 µA ± 14 µA ± 14 µA
100 mA
[1]
1 kHz 100 mA Q ± 600 µA ± 600 µA ± 600 µA
10 mA 1 kHz 1 A ± 1.41 mA ± 1.41 mA ± 1.41 mA
1A
[1]
1 kHz 1 A ± 1.4 mA ± 1.4 mA ± 1.4 mA