User`s guide

82
Chapter 3 System Overview
Measurement Input
Totalizer The multifunction module has a 26-bit totalizer which can
count pulses at a 100 kHz rate. You can manually read the totalizer
count or you can configure a scan to read the count.
You can configure the totalizer to count on the rising edge or falling
edge of the input signal.
The maximum count is 67,108,863 (2
26
- 1). The count rolls over to “0”
after reaching the maximum allowed value.
You can configure the totalizer to read without affecting the count or
reset the count to zero without losing any counts.
Channel 03
+IN
-IN
Gate
26 Bits
Tota liz e
Gate
34970A Refresh UG.book Page 82 Wednesday, February 17, 2010 12:34 PM