User`s guide

271
Chapter 7 Tutorial
Measurement Fundamentals
4
7
To make a more accurate measurement, you should extend the copper
test leads of the internal DMM closer to the measurement and hold the
connections to the thermocouple at the same temperature.
This circuit will give accurate temperature measurements. However, it is
not very convenient to make two thermocouple connections and keep all
connections at a known temperature. The Law of Intermediate Metals
eliminates the need for the extra connection. This empirical law states
that a third metal (iron (Fe) in this example) inserted between two
dissimilar metals will have no effect upon the output voltage, provided
the junctions formed are at the same temperature. Removing the
reference thermocouple makes the connections much easier.
This circuit is the best solution for accurate thermocouple connections.
Measurement
Thermocouple
Reference Thermocouple
Ice Bath
Internal DMM
Internal DMM
Measurement
Thermocouple
Ice Bath
(External Reference Junction)