User`s guide
27
Chapter 1 Quick Start
To Connect Wiring to a Module
4
1
To Connect Wiring to a Module
20 AWG Typical
6 mm
2 Connect wiring to the screw terminals.
.
Slot Channel
Channel Number:
1 Remove the module cover.
3 Route wiring through strain relief.
5 Install the module into mainframe.
Cable Tie Wrap
(optional)
• For detailed information on each module, refer
to the section starting on page 200.
• To reduce wear on the internal DMM relays,
wire like functions on adjacent channels.
• For information on grounding and shielding,
see page 259.
• The diagrams on page 28 show how to connect
wiring to a multiplexer module for each
measurement function.
4 Replace the module cover.
Wiring Hints...
34970A Refresh UG.book Page 27 Wednesday, February 17, 2010 12:34 PM