User`s guide

231
Chapter 5 Error Messages
Instrument Errors
4
5
450 Overrun during data collection: readings lost in USB transfer
Internal error: readings were collected too fast and were not buffered for
output to the USB drive.
451 Overrun during USB output: readings lost in USB transfer
Internal error: USB write operation was unable to keep up with data
collection.
452 Reading memory export aborted due to measurement reconfig
The export of reading memory was aborted because of measurement
reconfiguration.
453 Not able to execute while logging data to USB
Operation can not be completed while data is being actively logged to
USB.
454 Not able to execute while copying data to USB
Operation can not be completed while data is being exported to USB.
455 Not able to execute while importing a configuration from USB
Operation can not be completed while a measurement configuration is
being imported from USB.
457 Logging request ignored: USB device is busy
Logging was not started because USB is busy, but the scan will continue
to run normally, placing data in reading memory.
458 External USB drive is inaccessible
External USB drive can not be accessed; either the disk is full or else it
may need to be reformatted. The instrument will behave as if no drive is
present. The instrument was unable to find a valid partition to use for
storing instrument data.
34970A Refresh UG.book Page 231 Wednesday, February 17, 2010 12:34 PM