Chapter 5 Error Messages
Instrument Errors
Instrument Errors
111 Channel list: slot number out of range
The specified slot number is invalid. The channel number has the form
(@scc), where s is the slot number (100, 200, or 300) and cc is the
channel number. Example: CONF:VOLT:DC (@404)
112 Channel list: channel number out of range
The specified channel number is invalid for the module in the selected
slot. The channel number has the form (@scc), where s is the slot
number (100, 200, or 300) and cc is the channel number.
Example: ROUT:CLOSE (@134)
113 Channel list: empty scan list
Before you can initiate a scan, you must set up a scan list which includes
all configured multiplexer or digital channels in the instrument. Use the
MEASure?, CONFigure, or ROUTe:SCAN commands to set up your scan
list.
201 Memory lost: stored state
This error is reported at power-on to indicate that a stored state has
become unusable. This error is most likely caused by a dead battery
(memory is battery-backed). Refer to the 34970A/34972A Service Guide
to replace the internal battery.
202 Memory lost: power-on state
This error is reported at power-on to indicate that the power-down state
of the instrument (normally recalled when power is turned on) has
become unusable. This error is most likely caused by a dead battery
(memory is battery-backed). Refer to the 34970A/34972A Service Guide
to replace the internal battery.