Chapter 5 Error Messages
Execution Errors
-214 Trigger deadlock
A trigger deadlock occurs when the trigger source is “BUS” and a READ?
command is received.
-221 Settings conflict
An invalid configuration was requested. This error is most commonly
generated when setting alarm limits. Note that the lower limit must
always be less than or equal to the upper limit, even if you are using only
one of the limits. This error is also generated if you send the MEASure?
or CONFigure command with autorange enabled with a fixed resolution.
-222 Data out of range
A numeric parameter value is outside the valid range for this command.
Example: TRIG:COUNT -3
-223 Too much data
A character string was received but could not be executed because the
string length was more than 12 characters. This error can be generated
by the CAL:STRing and DISPlay:TEXT commands.
-224 Illegal parameter value
A discrete parameter was received which was not a valid choice for this
command. You may have used an invalid parameter choice.
Example: TRIG:SOURCE ALARM (ALARM is not a valid choice)
-230 Data stale
A FETCh? or DATA:REMove? command was received but internal
reading memory was empty. The readings retrieved may be invalid.
-310 System error
A firmware defect has been found. This is not a fatal error but you should
contact your nearest Agilent Service Center if this error is reported.