Chapter 4 Features and Functions
34908A 40-Channel Single-Ended Multiplexer
4
34908A 40-Channel Single-Ended Multiplexer
The module is divided into two banks of 20 channels each. All of the 40
channels switch HI only, with a common LO for the module. The module
has a built-in thermocouple reference junction to minimize errors due to
thermal gradients when measuring thermocouples.
NOTES:
Refer to the diagrams on page 27 to connect wiring to the module.
Only one channel can be closed at a time; closing one channel will open the previously
closed channel.
This module cannot be used to directly measure current or any 4-wire measurements.
When connecting thermocouples to the screw terminals on this module (not recommended
due to the common LO configuration), be sure to provide electrical isolation between
thermocouples to avoid current loops and subsequent measurement errors.
Connections to AC line are not recommended unless you provide external transient
Maximum Input Voltage: 300 V (CAT I)
Maximum Input Current: 1 A
Maximum Switching Power: 50 W
WARNING: To prevent electrical shock, use only wire that is rated for the highest voltage
applied to any channel. Before removing a module cover, turn off all power to external devices
connected to the module.
Bank Switch
Channel Switches
Backplane
Internal
DMM Input
Reference
Junction
Sensor
01
20
Com
21
40
98
99
Switch
Com
6mm
20 AWG Typical