User`s guide
182
Chapter 4 Features and Functions
USB Drive Front Panel - 34972A
Formatting Readings
You can control how readings are formatted on the USB drive. In
particular, you can choose whether the readings are stored in one large
file (
ROWS/FILE:AUTO)
or in a series of files with 64K - 1 rows per file
(
ROWS/FILE:64K)
. You can also choose whether the files are formatted
with a tab, comma or semicolon between columns.
• Front-Panel Operation:
FORMAT READNG ROWS/FILE:AUTO
FORMAT READNG ROWS/FILE:64K
FORMAT READNG SEP: TAB
FORMAT READNG SEP: COMMA
FORMAT READNG SEP: SEMICOLON
Importing an Instrument Configuration
You can import an instrument configuration stored in an Agilent
BenchLink Data Logger configuration (BLCFG) file in the root directory
of your USB drive.
• Front-Panel Operation:
IMPORT CONFIG
34970A Refresh UG.book Page 182 Wednesday, February 17, 2010 12:34 PM