User`s guide

167
Chapter 4 Features and Functions
System-Related Operations
4
Internal DMM Disable
You can scan through the configured channels using either the internal
DMM or an external instrument. For externally-controlled scans, you
must either remove the internal DMM from the instrument or
For information on controlling a scan with an external instrument,
refer to “Scanning With External Instruments” on page 111.
When shipped from the factory, the internal DMM is enabled. When
you change the state of the internal DMM, the instrument issues a
Factory Reset (*RST command).
A Factory Reset (*RST command) or Instrument Preset (SYSTem:
PRESet command) does not affect the internal DMM configuration.
Front-Panel Operation:
DMM ENABLED , DMM DISABLED
Remote Interface Operation:
INSTrument:DMM {OFF|ON}
Firmware Revision Query
The instrument has three microprocessors for control of various internal
systems. Each plug-in module also has its own on-board microprocessor.
You can query the instrument and each module to determine which
revision of firmware is installed for each microprocessor.
The instrument returns three revision numbers. The first number is
the firmware revision number for the measurement processor; the
second is the input/output processor; and the third is the front-panel
display processor. For each plug-in module, the instrument returns
one revision number for the on-board processor.