User`s guide
167
Chapter 4 Features and Functions
System-Related Operations
4
Internal DMM Disable
You can scan through the configured channels using either the internal
DMM or an external instrument. For externally-controlled scans, you
must either remove the internal DMM from the instrument or
• For information on controlling a scan with an external instrument,
refer to “Scanning With External Instruments” on page 111.
• When shipped from the factory, the internal DMM is enabled. When
you change the state of the internal DMM, the instrument issues a
Factory Reset (*RST command).
• A Factory Reset (*RST command) or Instrument Preset (SYSTem:
PRESet command) does not affect the internal DMM configuration.
• Front-Panel Operation:
DMM ENABLED , DMM DISABLED
• Remote Interface Operation:
INSTrument:DMM {OFF|ON}
Firmware Revision Query
The instrument has three microprocessors for control of various internal
systems. Each plug-in module also has its own on-board microprocessor.
You can query the instrument and each module to determine which
revision of firmware is installed for each microprocessor.
• The instrument returns three revision numbers. The first number is
the firmware revision number for the measurement processor; the
second is the input/output processor; and the third is the front-panel
display processor. For each plug-in module, the instrument returns
one revision number for the on-board processor.