Remote Interface Operation: The following command turns off the
front panel display.
DISPLAY OFF
The following command displays a message on the front panel and
turns on the display if disabled.
DISP:TEXT ’SCANNING ...’
To clear the message displayed on the front panel (without changing
the display state), send the following command.
DISPLAY:TEXT:CLEAR
Real-Time System Clock
During a scan, the instrument stores all readings and alarms with the
current time and date. The instrument stores the time and date
information in non-volatile memory.
When shipped from the factory, the instrument is set to the current
time and date (U.S. Mountain Time).
Front-Panel Operation:
TIME 03:45 PM
J
UN 01 1997
Remote Interface Operation: Use the following commands to set the
time and date.
SYST:TIME 15,45,00
Set time to 3:45 PM
SYST:DATE 1997,06,01
Set date to June 1, 1997