Chapter 4 Features and Functions
Totalizer Operations
4
Totalizer Operations
The multifunction module has a 26-bit totalizer which can count TTL
pulses at a 100 kHz rate. You can manually read the totalizer count or
you can configure a scan to read the count.
The totalizer channel is numbered “s03”, where s represents the slot
number.
You can configure the instrument to count on the rising edge or
falling edge of the input signal.
You can control when the totalizer actually records counts by
providing a gate signal
(G and terminals on the module). A TTL
high signal applied to the “G” terminal enables counting and a low
signal disables counting. A TTL low signal applied to the “
terminal enables counting and a high signal disables counting. The
totalizer only counts when both terminals are enabled. You can use
either the G terminal, the terminal, or both. When a gate is not
connected, the gate terminal floats to the enabled state, effectively
creating a “gate always” condition.
G
G
G
Input Signal
(Rising Edge)
Gate Signal
Total izer I nput
(High True)
Add to Total