User`s guide

138
Chapter 4 Features and Functions
Mx+B Scaling
The maximum gain allowed is ±1E+15 and the maximum offset
allowed is ±1E+15.
•The MEASure? and CONFigure commands automatically set the gain
(“M”) to 1 and offset (“B”) to 0.
A Factory Reset (*RST command) turns off scaling and clears the
scaling values on all channels. An Instrument Preset
(SYSTem:PRESet command) or Card Reset (SYSTem:CPON command)
does not clear the scaling values and does not turn off scaling.
Front-Panel Operation: The menu automatically guides you through
the gain, offset, and measurement label settings.
SET GAIN , SET OFFSET , SET LABEL
To reset the gain, offset, and measurement label to their defaults, go
to the corresponding level in the menu and turn the knob. To turn
scaling off (without clearing the gain and offset values), go to the first
level in the menu and select SCALING OFF.
SET GAIN TO 1 , SET OFST TO 0 , DEFAULT LABEL
To make a null measurement and store it as the offset, go to
SET OFFSET in the menu and turn the knob.
SET OFFSET
Remote Interface Operation: Use the following commands to set the
gain, offset, and custom measurement label.
CALC:SCALE:GAIN 1.2,(@101)
CALC:SCALE:OFFSET 10,(@101)
CALC:SCALE:UNIT ’PSI’,(@101)
After setting the gain and offset values, send the following command
to enable the scaling function on the specified channel.
CALC:SCALE:STATE ON,(@101)
To make a null measurement and store it as the offset, send the
following command.
CALC:SCALE:OFFSET:NULL (@101)