User`s guide

116
Chapter 4 Features and Functions
General Measurement Configuration
For frequency and period measurements, the "range" parameter is
used to compute a specific measurement resolution (see the Agilent
34970A/34972A Programmer’s Reference Help for details). When
specifying a (non-default) resolution, both the range and resolution
parameters must be specified within the MEASure? and CONFigure
commands. Refer to the Agilent 34970A/34972A Programmer’s
Reference Help for more information.
•The MEASure? and CONFigure commands contain an optional range
parameter which allows you to specify the range or autoranging.
The instrument returns to autoranging when the measurement
function is changed and after a Factory Reset (*RST command). An
Instrument Preset (SYSTem:PRESet command) or Card Reset
(SYSTem:CPON command) does not change the range setting.
Front-Panel Operation: First, select the measurement function on the
active channel. You are automatically guided to the next level of the
menu where you can select a specific range or autorange.
100 mV RANGE
Remote Interface Operation: You can select the range using
parameters in the MEASure? and CONFigure commands. For
example, the following statement selects the 10 VDC range on
channel 301.
CONF:VOLT:DC 10,DEF,(@301)