Specifications

Plug-in Module Test Considerations
For optimum performance, all test procedures should comply with the
following recommendations:
Assure that the calibration ambient temperature is stable and
between 18 °C and 28 °C. Ideally the calibration should be
performed at 23 °C ± 1 °C.
Assure ambient relative humidity is less than 80%.
Install the plug-in module and allow a 45 minute warm-up period
before verification or adjustment.
Use shielded twisted pair PTFE insulated cables to reduce settling
and noise errors. Keep the input cables as short as possible.
Remove all user wiring and connections from the plug-in modules
before verification or adjustment.
Use 4-wire Ohms measurement techniques for checking relay
contact resistance. Check directly at the terminals where possible.
Module Reference
4
Chapter 4 Calibration Procedures
Plug-in Module Test Considerations