Specifications

Calibration Procedures
This chapter contains procedures for verification of the instrument’s
performance and adjustment (calibration). These procedures are
required only if the internal DMM is installed. The chapter is divided
into the following sections:
Keysight Technologies Calibration Services, on page 63
Calibration Interval, on page 63
Time Required for Calibration, on page 64
Automating Calibration Procedures, on page 64
Recommended Test Equipment, on page 65
Input connections, on page 66
Calibration Security, on page 67
Calibration Message, on page 69
Calibration Count, on page 69
Calibration Procedures, on page 70
Aborting a Calibration in Progress, on page 70
Test Considerations, on page 71
Performance Verification Tests, on page 72
Internal DMM Verification Tests, on page 75
Optional AC Performance Verification Tests, on page 80
Internal DMM Adjustments, on page 81
–10 Vdc Adjustment Procedure (Optional), on page 85
Plug-in Module Test Considerations, on page 87
Relay Verification, on page 88
Thermocouple Reference Junction (Optional), on page 112
34907A Analog Output, on page 114
Closed-Case Electronic Calibration The instrument features
closed-case electronic calibration. No internal mechanical
adjustments are required. The instrument calculates correction
factors based upon the input reference value you set. The new
correction factors are stored in non-volatile memory until the next
calibration adjustment is performed. Non-volatile EEPROM
calibration memory does not change when power has been off or
after a remote interface reset.
62